Main Article Content
Abstract
Introduction. Hepatic micrometastases smaller than 10 mm critically influence oncologic staging and eligibility for curative metastasectomy, yet the sub-centimetre sensitivity of standard gadolinium-enhanced MRI is limited. High-relaxivity macrocyclic gadolinium-based contrast agents (GBCAs) generate higher lesion-to-liver contrast-to-noise ratio (CNR) and may improve detection. We compared a high-relaxivity versus a standard macrocyclic GBCA at equimolar dose within the same patients.
Methods. In this prospective intra-individual diagnostic-accuracy study reported per STARD 2015 at a tertiary hospital in Palembang, Indonesia, 48 adult oncology patients underwent two 3.0-T liver MRI examinations 2 to 7 days apart in randomised contrast order (standard gadoterate meglumine versus a high-relaxivity macrocyclic agent, 0.1 mmol/kg). Two gastrointestinal radiologists blinded to agent and clinical and reference data read images independently. A composite reference standard (histopathology and ≥6-month multiphasic imaging follow-up) defined 134 metastatic lesions. Sensitivity, specificity, AUC (DeLong), likelihood ratios, Cohen kappa and McNemar tests were computed.
Results. The high-relaxivity agent increased CNR by 12.2 units (48.5 versus 36.2; p<0.001). Per-lesion sensitivity rose to 94.8% (95% CI 89.6–97.4) from 82.8% (75.6–88.3), and micrometastasis sensitivity to 88.7% from 66.1% (McNemar p<0.001). The area under the ROC curve was 0.929 (0.893–0.966) versus 0.821 (0.757–0.886; DeLong p=0.003), and the negative likelihood ratio improved to 0.06. Inter-reader kappa was 0.89 versus 0.83. Specificity was comparable (84.5% versus 89.7%; p=0.41).
Conclusion. At equimolar dose, the high-relaxivity macrocyclic GBCA significantly improved CNR and sub-centimetre hepatic metastasis detection without meaningful loss of specificity, a gain that may alter oncologic management. High-relaxivity agents are recommended for high-risk hepatic staging.
Keywords
Article Details
Sriwijaya Journal of Radiology and Imaging Research (SJRIR) allow the author(s) to hold the copyright without restrictions and allow the author(s) to retain publishing rights without restrictions, also the owner of the commercial rights to the article is the author.